Questions related to use of OFM push-pull pumps
Yes, demo pumps are available upon request through BASi.
Fri, 3 Aug, 2018 at 12:04 PM
Yes, the OFM pump (MPP102PC) has push and pull options in the same unit. Either one pump head can be used for push and the second pump head for pull or, wit...
Fri, 3 Aug, 2018 at 12:19 PM
Flow rates can be adjusted between 0.1–10 µl/min with an accuracy of ± 20%. We recommend using pre-calibrated tubing to ensure sufficient accuracy. Furtherm...
Fri, 3 Aug, 2018 at 12:22 PM
Lowest flow rate: 0.1 µl/min Highest flow rate: 10 µl/min. Flow rates can be adjusted between 0.1 – 10 µl/min with an accuracy of ± 20%. We recommend us...
Fri, 3 Aug, 2018 at 12:23 PM
For better sampling results, we recommend using the OFM perfusate bag, which is delivered fully sealed in gamma irradiated protective packaging. However, if...
Fri, 3 Aug, 2018 at 12:25 PM
Joanneum Reaserach is considering a CPU control for the next version but unfortunately it is not yet available for the current version.
Fri, 3 Aug, 2018 at 2:25 PM
Please refer to the OFM pump user manual which is also available on the BASi webpage (Resources & Support). Chapter “Installing Pump Tubing and Perfusat...
Mon, 29 Oct, 2018 at 3:36 PM
There are tubing sets available on the BASi webpage that support operating up to 3 probes in parallel (e.g. OFM-P3-110, Three Channel Push Tubing or OFM-PH3...
Mon, 29 Oct, 2018 at 3:38 PM
Flush mode is generally used to flush the tubing set and get rid of air bubbles. Thus the pump is operated at a freely selectable flowrate of e.g. 10 µl/min...
Mon, 29 Oct, 2018 at 3:45 PM
cOFM probes are normally used at the same flow rate for P1 and P2 (e.g. 0.3 to 1 µl/min), but as the pump is also designed to operate dermal probes, flow ra...
Mon, 29 Oct, 2018 at 3:48 PM